A&D BA-T Series Instruction Manual page 36

Analytical balance
Hide thumbs Also See for BA-T Series:
Table of Contents

Advertisement

Counting mode sample input
Display settings: HOME key
Application selection button
HOME key
Input by weighing samples button
1
Sample input mode display
2
Number of samples input button
3
Back button
4
SAVE button
Name
Sample input mode
1
display
Number of samples
2
input button
3
Back button
4
SAVE button
This screen is used to input the unit weight by weighing samples for the Counting mode.
How to store the unit weight
Counting example for BA-6DTE
Step 1.
Press the Number of samples input
button.
*
Note that a greater number of sample
pieces will yield more accurate counting
result since the sample unit weight is
usually considered to vary more or less.
Step 2.
Place a container on the weighing pan
and press the RE-ZERO button
set the weighing value to zero.
Step 3.
Place the specified number of sample
pieces on the container/weighing pan.
Step 4.
Press the SAVE button
unit weight has been stored." is displayed,
the storage is complete.
Press the HOME key
Counting mode screen.
Caution
If the balance judges that the unit weight is too light to be stored, the SAVE button
The stored unit weight is stored in nonvolatile memory even if the power is removed.
APP button
[Counting mode] screen
Displayed in the Counting mode sample input mode.
Input the number of samples for the input by weighing samples.
Input range: 10 - 10000 pcs
(10 - 100 pcs with the touch screen software version 1.020 or earlier)
Displays the [Counting mode] screen.
Store the unit weight based on the current weight value and number of samples.
to
. When "The
to return to the
[Sample input mode] screen
[Application] settings screen
Select [Counting mode].
Unit weight setting button
Counting mode [Sample input mode] screen
Description
Number of samples input button
Numerical value input screen for
inputting the number of samples
Container (tare)
RE-ZERO button
Weighing with the specified number of samples
SAVE button
Counting mode display
36
Zero display
Storage complete
is disabled.

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ba-6teBa-6dteBa-225teBa-225dteBa-125dte

Table of Contents