Commentary on filter settings
Response characteristics
The display shows an acute response to weight variation. Reduce the setting value when
FAST
weighing powder or liquid, weighing extremely lightweight samples, or prioritizing the
operating efficiency over the stability of the weighing value.
The display shows a gradual response to weight variation. Increase the setting value when
SLOW
the weighing value does not get stabilized depending on the usage environment.
Stability bandwidth
The setting to determine that the weighing value is stable. It shows the stabilization indicator when the
fluctuation range in a certain time of period gets lower than the setting value, and outputs the weighing
value based on the internal settings. This setting affects the automatic printing.
The minimum display to be shown is 1 d.
(Example) If No blank is selected with BA-6TE and 0.000001-g display is chosen, 0.000001 g is
shown as 1 d.
The stabilization indicator does not appear unless the weighing value is stable enough, and
±1d
it disappears when the weighing values changes even slightly. Reduce the setting value for
precise weighing.
The display does not easily respond to a slight weight variation. Increase the setting value
±3d
when the weighing value does not get stabilized depending on the usage environment.
Zero tracking
The function to automatically track the zero point and keep the zero display when the display shows
zero and the zero point slightly fluctuates depending on usage environment. A tracking level can be
selected from 3 levels. Increase the setting value if zero is not stabilized. Do not perform zero tracking
when the weighing value is slightly over d. d is the minimum unit of the display.
BA-6TE /
Setting
value
BA-6DTE
Off
±2 d/1 second
Normal
±5 d/0.5 seconds
Strong
Very
±10 d/0.2 seconds
strong
BA-225TE /
BA-225DTE /
BA-125DTE
Off
±1 d/1 second
±1.5 d/0.5 seconds
±1.5 d/0.2 seconds
The zero point is not tracked with zero tracking.
The zero point is normally tracked with zero
tracking.
The zero point is strongly tracked with zero
tracking.
The zero point is very strongly tracked with zero
tracking.