Measurement example
Select a recipe to perform measurement according to the recipe. The registration example below is described
here as an example.
Recipe name
Target value unit
Solution volume
Step 1.
On the [HPLC mode] screen, press the Go to the recipe list button to display the [Recipe list] screen.
Select the recipe to be used for the measurement from the list and press the Back button to display the
[HPLC mode] screen.
Re-zero will be applied automatically when the screen changes to the [HPLC mode] screen.
Sample 1 is selected here as an example.
Step 2.
To record the tare value:
Press the SAVE button
Tare operation will be applied automatically when the SAVE button
The weighing screen for the first sample is displayed.
If the target value plus tare value of the first sample exceeds the capacity, the SAVE button
disabled.
To skip recoridng of the tare value:
Press the Skip button
The weighing screen for the first sample is displayed.
If a warning is displayed at the top of the screen:
Check the recipe name or the target value and tolerance value of the registered sample.
Step 3. Weigh the displayed sample.
When the stability indicator lights up and the SAVE button
is enabled, press the button to save the weighing value.
Allowable range: |(Weighing value) - (Target value)|
(Tolerance)
If the allowable range is exceeded:
To start over from tare weighing with the current sample, press
the Back button.
Step 4.
Repeat the tare weighing and sample weighing in steps 2 and 3 for every sample registered in the recipe.
Step 5.
The measurement results screen will be displayed when measurement of all samples is completed. Check
and output the results. Then, press the Back button to display the [HPLC mode] screen.
PBS(-)
mmol/L
Disodium hydrogen phosphate
1 L
Potassium dihydrogen phosphate
Sodium chloride
Potassium chloride
when the weighing value is 0 g or greater and stable.
.
Sample name
71
Target value
Tolerance
10.000 mmol/L
10.000 %
2.000 mmol/L
10.000 %
137.000 mmol/L
10.000 %
2.680 mmol/L
10.000 %
is pressed
is