APPENDIX C: Displayed Error Codes ..................13 APPENDIX D: Warranty and Service Information................14 Introduction The SB-30/70/150 Digital Bench Scales are 44 requirements. To re-configure the scale, compact bench scales with display and five please refer to the “Setup Menu Mode” sec- function keyboard.
DC power input, a cali- bration switch, a remote display connector RS-232 ON/OFF DC POWER REMOTE (SB-30 & SB-70 only) and an ON/OFF rocker SWITCH JACK PORT DISPLAY switch. To install the scale, simply plug the...
Display Overview GROSS TARE ZERO STABLE SB-30/70/150 Display Area ANNUNCIATOR MEANING Better known as the “Center of Zero” annunciator, this light is active whenever the ZERO displayed weight is within ± 0.25 divisions of true zero. Indicates that the scale is displaying net weight.
Keyboard Functions ZERO TARE PRINT GROSS SB-30/70/150 Series Keyboard FUNCTION Sets scale to display "0" only when a) in Gross mode, b) not in overload and c) not in ZERO motion. Toggles between Gross and Net weight display if a tare has been established.
User Menu Mode SB-30/70/150 includes a User menu which contains the scale's serial communication parameters. It consists of 5 separate menu selections, each also with its own sub-menu of choices. To change a listed parameter, you must first enter the User Menu mode. Once there, four of the front panel keys become directional navigators to move around in the menus, and one key is used to save or SET the selections.
Key Setup Menu Mode SB-30/70/150 scale include a Setup menu which determines the scale’s display mode, allows scale calibration and also contains the scale's adjustable functional parameters. It consists of 5 separate menu selections, each also with its own sub-menu of choices.
Page 8
Setup Menu Mode / Continued To place the unit in Setup menu mode: 1. Turn the power off to the unit. 2. Locate the calibration switch on the scale’s back panel and toggle to the opposite position. 3. While holding down both the ZERO and PRINT keys, turn the power back on.
For the SB-30, the minimum test weight is 0.3 lb (0.15 kg) , and the recommended test weight is 20 lb (10 kg). For the SB-70, the minimum test weight is 0.7 lb (0.35 kg) , and the recommended test weight is 47 lb (23 kg).
Page 10
Calibration / Continued To calibrate the scale: 1. If you are already in Setup Menu Mode, scroll to “F 12” and press the ZERO key. Otherwise, turn the scale off and set the Calibration Switch on the back panel to the opposite position. Turn the unit back on.
OPERATING TEMPERATURE RANGE: Base: Steel and ABS 32°F to 104°F (0°C to 40°C) Platform: Stainless Steel Feet: Non-skid Hard Rubber PLATTER SIZE: SB-30/70: 10” x 13” DISPLAY: SB-150: 13” x 13” 6 Digit, 0.6", 7-Segment Green LED POWER SOURCE: OVER CAPACITY ANNUNCIATION:...
APPENDIX B: Troubleshooting the Serial Port The scale’s serial port operation can be verified by following the procedure below. This procedure makes use of HyperTerminal - a serial port communication program bundled with Windows 95 and Windows 98 operating systems. A HyperTerminal configuration file is available on the Tran- scell website.
Internal resolution is not high enough to process keyed-in Err 2 weight value in Calibration Mode. Verify test weight and value. All Modes Diagnostics check error - EEPROM Read Err 3 Err 4 All Modes Diagnostics check error - EEPROM Write SB-30/70/150 Operation Manual...
APPENDIX D: Warranty and Service Information Seller warrants that each SB-30/70/150 will conform to written specifications, drawings, and other descriptions made by the manufacturer, including any modifications thereof. Seller warrants the goods against faulty workmanship and defective materials. If any goods fail to...