Solar 7000 Display Check

Revision J
MAINTENANCE: Checkout Procedure
4. Check all functions of the four control keys as follows.
If NBP is available, the NBP key will start the non-invasive blood
pressure pump. The pump may be stopped by selecting the key again
if no hose is connected. If NBP is not available, a message will be
displayed to state that no NBP is available.
ZERO ALL key will give a message that a BP parameter is needed.
SILENCE ALARM can be checked by disconnecting the patient cable
from the Tram module. Select the key again to silence the alarm.
GRAPH GO/STOP will start the graph or display a NO GRAPH
message.
1. For color Solar 7000 display only, degauss the monitor by executing
the following menu selections from the MAIN MENU.
MONITOR SETUP
SERVICE MODE
Password
Enter day and month from monitor screen with leading zeros.
(July 4 = 0407)
DEGAUSS
Degauss the color display periodically and whenever the red color
becomes impure (not true red). Note that the display is automatically
degaussed every time the unit is turned ON. Once the display is
manually degaussed, it cannot be degaussed again for at least 10
minutes.
2. Display FOCUS TEST PATTERN on the display by executing the
following menu selections from the MAIN MENU.
MONITOR SETUP
SERVICE MODE
Password
Enter day and month from monitor screen with leading zeros.
(July 4 = 0407)
HARDWARE TESTS
VIDEO TEST
YES
FOCUS TEST PATTERN
If the Solar 7000 display does not have clarity of line definition,
perform "Display Calibration" in Chapter 5: "Calibration."
Press Trim Knob control to exit.
Solar 7000/8000/View Patient Monitor
414993-001
3-17