Solar 7000 Disassembly Guidelines; General; Pcb Assemblies; Hardware - GE Solar 7000 Service Manual

Patient monitor
Table of Contents

Advertisement

UPPER LEVEL ASSEMBLY: Solar 7000 Disassembly Guidelines

Solar 7000 Disassembly Guidelines

General

PCB Assemblies

Hardware

Revision J
Two versions of the Solar 7000 monitor are available: color or
monochrome. Follow the guidelines described when disassembling your
version monitor. You will require a standard set of hand tools.
Observe the following guidelines when working on all PCB assemblies:
Take precautions against electrostatic discharge damage.
Handle all PCB assemblies by their edges.
Generally, use 60/40 low melting point solder. Use high melting point
silver solder when working on the surface mounted components.
Use a heatsink when soldering any phototransistors, photodiodes, or
optical isolators.
Repair breaks in PCB tracks by bridging with wire from pad to pad.
Do not lay solder across the breaks. (The heat from the soldering iron
may cause the track to detach from the PCB.)
After soldering, clean excess flux from the PCB assembly.
If you replace the main processor, remove the serial EEPROM U9
and insert it into the replacement main processor PCB. The serial
EEPROM must stay with the unit.
Before disassembly, note the positions of any wires or looms (cables),
marking them if necessary to ensure that they are replaced correctly.
Gray ribbon cables have retainer clips holding them in the connector.
Save and set aside all hardware for re-assembly.
Solar 7000/8000/View Patient Monitor
414993-001
7-3

Advertisement

Table of Contents
loading

This manual is also suitable for:

Solar 8000Solar solarview

Table of Contents