UPPER LEVEL ASSEMBLY: Solar 8000/View Disassembly Guidelines
Power Supply Removal
Revision J
Do the following to remove the power supply.
WARNING
Beware of AC line voltage at the power module. When
rear power switch is OFF, AC line voltage is still present.
Capacitors C1-C4 may remain fully charged up to 30
seconds. Use an insulated screw driver at all times when
making adjustments.
1. Verify power cord is disconnected.
2. Remove one end of the harness from the power supply assembly to
the main processor PCB.
CAUTION
Keep circuit board mounted to metal sub-chassis at all
times to avoid stressing solder joints and surface mount
components.
3. Remove four recessed screws attaching the power supply assembly.
Solar 7000/8000/View Patient Monitor
414993-001
7-25