TROUBLESHOOTING: Introduction
Configured memory contains the Ethernet address, internet address, bed
name, care unit name, and standard unit defaults for each monitor.
CONTINUE is a menu selection on the monitor that appears after a
successful software download. It allows the user to continue downloading
other files without resetting the monitor.
This front panel switch on the monitor only turns the display ON and
OFF. It has no effect on the software update, but it is used as a screen
saver.
Protected memory contains a patient's history and any individualized
changes to the unit defaults.
To POWER CYCLE or reboot the monitor, turn the power OFF at the
rear of the unit, not at the front switch. A POWER CYCLE initializes the
new software and updates the software revision window.
There are two service menus available. The service menu found in the
BOOT LOADER is used when downloading the boot code and main
processor code. The SERVICE MODE menu is imbedded in the monitor
MAIN menu and is used for various functions like calibration, video
tests, and downloading monitor interface software.
WARM START is a menu selection on the monitor that appears after a
successful software download. A WARM START will activate the
software previously downloaded. The following also activates a warm
start.
Hold down the NBP GO/STOP and ZERO ALL keys on the front
panel. (For the SolarView remote display controller, hold down the
two blank keys.)
Press and quickly release the Trim Knob control, NBP GO/STOP and
ZERO ALL keys.
Solar 7000/8000/View Patient Monitor
414993-001
4-5