Copying Stored Monitor
Defaults
Revision J
CONFIGURATION: Advanced User Procedures
WARNING
Copying Monitor Defaults from another monitor will
erase all of the Monitor Defaults on the current monitor.
To copy Monitor Defaults from the server monitor, the client monitor
must first be in the DISCHARGED state. If the client monitor is set to
the OR patient-monitor type, it must NOT be in Combo or Rover Combo
admit mode.
NOTE
Make sure that the client monitor shares the same
configuration as the server monitor.
1. At the client monitor, execute the following menu sequence, starting
from the MAIN menu select:
MONITOR SETUP
SERVICE MODE
Password
Enter day and month from monitor screen with leading zeros.
(July 4 = 0407)
2. From the Service menu select COPY UNIT DEFAULTS.
3. Select the SELECT UNIT TO COPY MONITOR DEFAULTS FROM
option to scroll through a list of units currently on the UNITY
network. Choose the unit that has the desired server monitor.
4. Select the SELECT BED TO COPY MONITOR DEFAULTS FROM
option to scroll through a list of beds within the selected unit. When
the desired bed is found, select it, and answer YES to the
confirmation popup menu.
NOTE
After transferring Monitor Defaults, the first Defaults set
is automatically activated. If another set of Defaults is
desired, the user must manually select it from the
RECALL DEFAULTS menu.
After copying Monitor Defaults from another bed, verify that the defaults
were transferred and arrhythmia levels are as desired.