UPPER LEVEL ASSEMBLY: Solar 7000 Disassembly Guidelines
Opening the Unit for
Service

Disassembly

7-4
Use these step-by-step disassembly instructions when you need to
remove various PCB assemblies from the monitor assembly.
WARNING
Make sure the monitor is not monitoring a patient,
because monitoring will be interrupted.
First, turn the unit OFF at the rear power switch and disconnect the AC
power cord and all communication cables.
WARNING
Due to possible high voltage present, use an insulated
screwdriver at all times when making adjustments.
1. Lay the unit on its front display on either foam or other soft material.
Do not apply excessive pressure on the Trim Knob control.
2. Remove the five screws at the rear as shown.
3. Lift the bezel directly up.
Solar 7000/8000/View Patient Monitor
414993-001
Revision J