Solar 7000 Color Power Supply

Preparation

5-10
CALIBRATION: Solar 7000 Color Power Supply
This procedure describes how to check and adjust the low-voltage power
supply PCB, pn 800974-00X, used in the Solar 7000 color monitor.
To perform this procedure, you will need the special tools and items listed
below. Equivalent equipment may be substituted.
Required Tools/Special Equipment
Item
Very small Blade Screwdriver
Multimeter (DMM)
WARNING
Beware of AC line voltage at the power module. When
rear power switch is OFF, AC line voltage is still present.
Use an insulated screw driver at all times when making
adjustments.
WARNING
Make sure the monitor is not monitoring a patient,
because monitoring will be interrupted.
Remove monitor rear power supply panel by following the steps given
below.
1. Turn power switch at the rear of the monitor to OFF.
2. Remove six screws attaching the rear panel to the power supply at
the rear of the unit. Beware of AC line voltage at the power module.
3. Carefully pull the heatsink/PCB out to make the component side of
the PCB accessible. Make sure the four cable connectors remain
intact.
Solar 7000/8000/View Patient Monitor
414993-001
Manufacturer and
Part Number/Model
Fluke 8060A