Color Crt Controller Pcb Removal - GE Solar 7000 Service Manual

Patient monitor
Table of Contents

Advertisement

UPPER LEVEL ASSEMBLY: Solar 7000 Disassembly Guidelines
Color CRT Controller PCB
Removal
7-8
For color monitors only, do the following.
WARNING
To avoid electrical shock on color units, do not touch the
large heatsink connected to the nearby deflection PCB,
pn 404642-00X, while the unit is ON.
Due to high voltage on either PCB, use an insulated
screw driver at all times while making adjustments.
1. Remove attaching harness.
2. Remove two screws and lift the color CRT controller PCB.
Solar 7000/8000/View Patient Monitor
414993-001
Revision J