What Error Data Is Useful - GE Solar 7000 Service Manual

Patient monitor
Table of Contents

Advertisement

What Error Data is
Useful?
6-22
CONFIGURATION: Advanced User Procedures
When troubleshooting a monitor, the parameters displayed below will be
of interest.
Name
Process Name
Error Code
Severity
Date
Time
Error Number
For some categories of Tram-net network errors, two additional lines are
added to the error log entry.
Name
Network Error
Channel Number
Solar 7000/8000/View Patient Monitor
414993-001
Definition of Parameters
Description
is the name of the software task that was
operating when the event/problem occurred.
is a code for the type of event/problem that
occurred.
indicates the impact of the event/problem on the
system.
is the date the event/problem occurred.
is the time the event/problem occurred.
is a sequential number that is used to identify
each event/problem.
Definition of Parameters
Description
identifies that a network error occurred.
identifies the network channel exhibiting the
error.
Revision J