Troubleshooting Monitor
Defaults Transfer
6-16
CONFIGURATION: Advanced User Procedures
Below is a list of error messages which may be displayed when
performing a Monitor Default Transfer.
Storing Monitor Defaults
Message
ERROR - UNIT DEFAULTS
HAVE NOT BEEN SAVED
Copying Stored Monitor Defaults
Message
ERROR COPYING UNIT
DEFAULTS - NETWORK
ERROR
ERROR COPYING UNIT
DEFAULTS - UNIT TYPE
MISMATCH
ERROR COPYING UNIT
DEFAULTS - SOFTWARE
VERSION MISMATCH
ERROR COPYING UNIT
DEFAULTS - DATA
CORRUPTION ERROR
ERROR COPYING UNIT
DEFAULTS - SOFTWARE
FEATURE MISMATCH
ERROR COPYING UNIT
DEFAULTS - COUNTRY CODE
MISMATCH
Solar 7000/8000/View Patient Monitor
414993-001
Description
Monitor Defaults were not saved
due to a memory error in the
monitor.
Description
Either an error has occurred while
the Defaults were being
transferred, or Defaults have not
been saved on the server monitor.
The server and client monitors are
set to different unit types (i.e.
ADULT, NEO, or OR)
The server and client monitors
have either different revisions of
software, or different software
versions (i.e. Solar 7/8000 vs.
Solar 8000M)
The Monitor Defaults transferred
in have become corrupted,
possibly due to a memory error.
The server and client monitors are
set to different software levels (i.e.
7020, 7025, etc.).
The server and client monitors are
set to different country codes (i.e.
DEFAULT, FRANCE, or
GERMANY).
Revision J
Need help?