4. After the parameter settings are complete, press [Enter].
5.7 Sense Function (Sense)
This menu item is used to switch the power supply to local measurement or re-
mote sensing.
Category
Parameter
Upper limit of voltage: Vlim
Upper limit of current: Ilim
Upper limit of power: Plim
[On/Off] status
Config
Mode
menu
Speed
V-Rise Time/I-Rise Time
V-Fall Time/I-Fall Time
Output Res
On Delay/Off Delay
•
Last: Indicates when powered on, the instrument will remain the same
parameter settings and output status as last time you powered off the
instrument.
•
Last+Off: Indicates when powered on, the instrument will remain the
same settings as last time you powered off the instrument, but the output
status is Off.
For example, Last is selected, and the voltage value is set to 20V. After the
instrument is powered off and then powered on, the voltage displayed on the
interface is 20V.
Copyright © Itech Electronic Co., Ltd.
Basic Operation
Initial Value
Upper limit value: One
percent of the rated
voltage of the
instrument
One percent of the
rated current of the
instrument
Rated power value of
the instrument
Off
CV
High
0.1s
1000
0
87