Teledyne API Model 401 O
Table 2-4: Warning Messages
Warning Message
SYSTEM RESET
RAM INITIALIZED
ANA LAMP WARNING
O3 LAMP SHUTDOWN
O3 FLOW WARNING
O3 REF DET WARN
SAMPLE FLOW WARNING
SAMPLE PRESSURE WARN
SAMPLE TEMP WARNING
BOX TEMP WARNING
O3 GEN LAMP TEMP WARN
ANA LAMP TEMP WARN
V/F NOT INSTALLED
The menu field changes depending on the mode of the Analyzer and the buttons that have been
pressed. It indicates the current function of each of the 8 pushbuttons below the display. See
Section 2.2.2 for information on using the pushbuttons.

2.2.2 Programmable Pushbuttons

The 8 pushbuttons below the display are programmable by the CPU in that their functions
change depending on the mode of the Analyzer or the operations being performed. The legend
above a button identifies its current function. If there is no legend above a button, it has no
function and will be ignored if pressed.
Photometric Calibrator Instruction Manual, 01124, Rev. J1
3
Meaning
Issued whenever Analyzer is powered on
RAM was erased (including DAS reports)
Analyzer lamp < 40000 OR >= 80000 counts
O
lamp temperature control not working
3
O
GEN flow < 1.8 L/M or > 5.5 L/M
3
O
reference detector < 50 mV
3
Sample flow < 500 cc/m or > 1000 cc/m
Sample pressure < 15 or > 35 In-Hg-A
Sample temperature < 12 or > 48 deg. C
Box temp. < 12 deg. C or > 48 deg. C
O
lamp < 43 or > 53 deg. C
3
Analyzer lamp < 51 or > 61 deg. C
A/D - I/O card not installed, bad, or no handshake.