Status Lines; Uv Lamp Power Supply Adjustment; Table 8-1: The Calibrator - Teledyne 401 Instruction Manual

O3 calibrator
Table of Contents

Advertisement

Teledyne API Model 401 O

8.7 Status Lines

Table 8-1: The Calibrator Has The Following Status Conditions
Pin Pair
Output #
High/Low
1
2
3
4
5
6
7
8
9
10

8.8 UV Lamp Power Supply Adjustment

Adjust the drive power of the lamp power supply as follows:
1. Remove the cover of the lamp power supply. Attach a DVM across TP7 and TP10, and
adjust the pot (VR1) until the DVM reads 20 volts ± 1 volt.
2. Adjust the positioning of the source lamp, as follows:
a. At the front panel of the instrument, Press the TEST key until O
displayed.
b. Loosen the lamp retaining thumb-screw and rotate the lamp until the O
display is 4500 mV ± 320 mV. Re-tighten the thumb-screw. (Note that the full range of
lamp adjustment can be achieved within ¼ revolution of the lamp. Note also that the O
REF display is updated approximately once every six seconds, and slow rotation of the
lamp is needed for proper adjustment.)
Photometric Calibrator Instruction Manual, 01124, Rev. J1
3
Status
1, 2
ZERO CAL
3, 4
SPAN CAL
5, 6
FLOW ALARM
7, 8
TEMP ALARM
9, 10
DIAG MODE
11,12
POWER
13,14
PRESS ALARM
15,16
IZS FLOW
17,18
SYSTEM OK
19,20
LAMP WARNING
Condition
In Zero Calibrate
In Span Calibrate
Flow Warning
Any Temp Warning
In Diagnostic Mode
On As Long As M401 Is Running
Alarms On Low Pressure
Warn Alarms If No IZS Flow
Alarms If Any Fault Is Present
Alarms If The Calibrator Lamp Intensity
Is Out Of Limits
8-7
REF=XXXXX is
3
REF reading on
3
3