Teledyne API Model 401 O
Table 7-6: Special Diagnostic Commands
Command
"D SYS-RESET<CRLF>"
"D RAM-RESET<CRLF>"
"D EE-RESET <CRLF>"
Whenever the diagnostic mode is entered or exited, a report is issued to the RS-232 output.
Table 7-7 summarizes the diagnostic reports.
Table 7-7: Diagnostic Reports
Report
"C DDD:HH:MM IIII ENTER DIAGNOSTIC MODE"
"C DDD:HH:MM IIII EXIT DIAGNOSTIC MODE"

7.5 Test Measurements

All the test measurements which can be displayed by pressing the TEST button are also available
to the host computer via the RS-232 interface. The host computer should issue a request for a test
measurement, and then the CPU will send the current value of the test measurement to the
RS-232 output. The format of the test measurement message is:
"T
DDD:HH:MM
For example, the format of the PMT output in millivolts would be:
"T
194:11:29
To request a test measurement, the host must issue a command of the form:
"T MEASUREMENT<CRLF>"
For a summary of all test functions issue the command "T LIST". Table 7-8 lists the commands
and the corresponding test measurements which will be returned.
Photometric Calibrator Instruction Manual, 01124, Rev. J1
3
Function
Reset Calibrator (same as power off/on)
Reset Calibrator and erase RAM
Reset Calibrator and erase RAM & EEPROM
IIII
TEST MEASUREMENT<CRLF>"
0000
PMT= 254 MV<CRLF>"
7-7