Table 2-2: System Modes - Teledyne 401 Instruction Manual

O3 calibrator
Table of Contents

Advertisement

Teledyne API Model 401 O

Table 2-2: System Modes

Mode
STBY xxxx (1)
ZERO CAL x (2)
O
GN CAL x (2)
3
ASEQ xxx (5) x (2)
HOLD OFF
D/A OUTPUT (3)
BCAL ZERO
BCAL SPAN
SETUP xxx (4)
(1) xxxx = AZS ( Auto Zero & Span programmed), AZ2S (Auto Zero & 2 span points; 25% &
100% of O
set concentration) programmed AZ5S
3
(Zero & 5 span points; 20%, 40%, 60%, 80% & 100% of O
(2) x = M (manual), A (auto), R (remote)
(3) diagnostic test modes
(4) xxx = software revision
(5) 0%, 20%, 25%, 40%, 60%, 80%, 100%
The message field shows test measurements or warning messages. Table 2-3 and Table 2-4
summarize the test measurements and warning messages and their meanings. Refer to Sections 4
and 5 for detailed information on viewing test measurements and warning messages and clearing
warnings.
Photometric Calibrator Instruction Manual, 01124, Rev. J1
3
Meaning
Standby mode (Photometer is operating normally)
Zero check in progress
Span check in progress
Multi-point calibration in progress
Hold off delay after power reset
Test analog output channels
Bench zero calibration
Bench span calibration
Configuring analyzer (operation continues)