Table 4-1: Setup Variables - Teledyne 401 Instruction Manual

O3 calibrator
Table of Contents

Advertisement

Teledyne API Model 401 O

Table 4-1: Setup Variables

Button Sequence
SETUP-EVNT-WAIT
SETUP-EVNT-TIME
SETUP-EVNT-ASEQ
SETUP-EVNT-SHFT
SETUP-COMM-BAUD
SETUP-COMM-ID
SETUP-MISC-O3-SLPE
SETUP-MISC-O3-GEN-ADJ
SETUP-MISC-O3-GEN-CAL
SETUP-MISC-O3-GEN-
MODE
SETUP-MISC-O3-OFFS
SETUP-MISC-DARK-VIEW
SETUP-MISC-CLK-TIME
SETUP-MISC-CLK-DATE
SETUP-MISC-CLK-ADJ
SETUP-MISC-D/A-RNGE
SETUP-MISC-D/A-OFFS
SETUP-MISC-MORE-FLT
SETUP-MISC-MORE-UNITS
SETUP-MISC-MORE-TCHN
SETUP-PASS
SETUP-CFG-LIST
Photometric Calibrator Instruction Manual, 01124, Rev. J1
3
Function
Zero cal. duration
Span cal. duration
Cal. time-of-day
Auto. Multipoint cal. enable
Auto check shift
RS-232 baud rate
Analyzer ID number
Slope adjust parameter
O
GEN lamp setup/adjust
3
O
GEN calibration
3
Feedback mode selection
Offset adjust parameter
Dark current offset
Current time-of-day
Current date
Clock speed adjust
D/A output range
D/A offset setting
Fault LED timeout
O
conc. units
3
TEST to analog output
Cal. password enable
Software configuration
4-7
Default
Limits
15 min.
1-20 min.
15 min.
1-20 min.
23:30
00:00-23:59
OFF
OFF/ZS/Z2S/Z5S
± 60 min.
0 min.
2400 baud
300, 1200, 2400
0000
0000-9999
1.00
0.85 - 1.15
OFF
ON - OFF
OFF
ON - OFF
BNCH
± 1000.0
0.0
0 mV
1000 mV
00:00
00:00 - 23:59
01 JAN 00
31 DEC 99
± 60 sec.
0 sec/day
500 ppb
100-10000
± 500 mV
0 mV
5 min
0 - 10 min
ppb
ppb/ppm/ug/mg
OFF
OFF-ON
ON
OFF-ON
O
CALIB.
Display