Table 9-3: V/F Boardj; Table 9-4: V/F Board Dip - Teledyne 401 Instruction Manual

O3 calibrator
Table of Contents

Advertisement

Teledyne API Model 401 O
2. Confirm that all jumpers on the V/F board are set properly, as follows:
Table 9-3: V/F Board Jumpers - Factory Settings
Jumper
Setting
B1
1
B2
1-2
B12
3-4 (0-5V)
B14
2-3 (Enabled)
B15
Set to match power line frequency
JP1
1-2
JP2
1-2
Table 9-4: V/F Board Dip Switches - Ranges For Analog Output
Switch
S1(DAC 0)= "REC"
S2(DAC 1)= "DAS"
S3(DAC 2)="TEST"
S3(DAC 2)="O3 GEN"
3. If Voltages and Jumper settings are correct, the V/F card is faulty and should be replaced.
Photometric Calibrator Instruction Manual, 01124, Rev. J1
3
Factory Set Jumpers
User Set DIP Switches
100 mV
Full Scale
1,6 = ON
1,6 = ON
1,6,7 = ON
N/A
1 V
5 V
Full Scale
Full Scale
1,5 = ON
1,4 = ON
1,5 = ON
1,4 = ON
1,5,7 = ON
1,4,7 = ON
N/A
1,4,7 = ON
9-12
10 V
Full Scale
1,3 = ON
1,3 = ON
1,3,7 = ON
N/A

Advertisement