Otto Bock 743A250 Bionic Pro Instructions For Use Manual page 50

Hide thumbs Also See for 743A250 Bionic Pro:
Table of Contents

Advertisement

Technical data
Surges
according
IEC 61000-4-5
Voltage drops, short inter­
ruptions and fluctuations of
the supply voltage accord­
ing to IEC 61000-4-11
Magnetic field with supply
frequency
(50/60 Hz)
according
IEC 61000-4-8
NOTE: U
is the mains voltage before applying the test levels.
T
Directives and manufacturer's declaration – electromagnetic interference immunity
The 743A250 Bionic Pro product is designed for operation in the electromagnetic environment specified below.
The customer or user of the product must ensure that it is operated in such an environment.
INTERFERENCE
IMMUNITY TESTS
Conducted
HF
interfer­
ence
according
IEC 61000-4-6
Radiated HF interference
according
IEC 61000-4-3
NOTE 1: The higher value applies at 80 MHz and 800 MHz.
50
to
±1 kV voltage
(line-to-line voltage)
0% U
; ½ period at 0, 45,
T
90, 135, 180, 225, 270 and
315 degrees
0% U
; 1 period
T
70% U
;
25/30
periods,
T
single phase: at 0 degrees
30 A/m
to
IEC 60601 test level
3 V
eff
to
150 kHz to 80 MHz
10 V/m
80 MHz to 2.7 GHz
to
6 V
in ISM frequency
eff
bands between 0.15 MHz
and 80 MHz 80% AM at
1 kHz
±1 kV voltage
(line-to-line voltage)
0% U
; ½ period at 0, 45,
T
90, 135, 180, 225, 270 and
315 degrees
0% U
; 1 period
T
70% U
;
25/30
periods,
T
single phase: at 0 degrees
30 A/m
Compliance level
3 V
eff
150 kHz to 80 MHz
10 V/m
80 MHz to 2.7 GHz
6 V
in ISM frequency
eff
bands between 0.15 MHz
and 80 MHz 80% AM at
1 kHz
The quality of the supply
voltage should be equival­
ent to a typical commercial
or hospital supply.
The quality of the supply
voltage should be equival­
ent to a typical commercial
or hospital supply.
Magnetic
fields
at
mains frequency should be
equivalent to the typical
levels for commercial or
hospital environments.
ELECTROMAGNETIC
ENVIRONMENT – direct­
ives
Portable and mobile wire­
less equipment should not
be used at a lesser dis­
tance from the product
including the lines than the
recommended safety dis­
tance calculated using the
equation applicable for the
transmission frequency.
Recommended
safety
distance:
d = 1.2√P
d = 1.2√P for 80 MHz to
800 MHz
d = 2.4√P for 800 MHz to
2.7 GHz
where P is the nominal out­
put of the transmitter in
watts (W) according to the
transmitter
manufacturer
and d is the recommended
safety distance in metres
(m). The field strength of
stationary radio transmit­
ters
at
all
frequencies
according to an on-site
investigation
a
should
less than the compliance
b
level.
Interference is possible in
the vicinity of devices that
bear the following symbol.
743A250 Bionic Pro
the
be

Advertisement

Table of Contents
loading

Table of Contents