8 Preparation for use

8.1 Alignment

The following alignment guidelines contain descriptions for connecting the knee joint to a prosthetic socket. In
principle, the alignment of the prosthesis is independent of the type of connection for the knee joint. In case of a
connection to an osseointegrated, percutaneous implant system, a socket is not used during bench alignment in
the alignment apparatus. In this case, the central proximal point on the prosthetic socket corresponds to the
trochanter of the thigh bone (see illustration in the section "Bench alignment in the alignment apparatus" see
Page 24).
Ensure that possible flexion or adduction of the transfemoral residual limb can be compensated to a permissible
extent by an adapter approved by the implant manufacturer in the course of static alignment optimisation. Safe
functioning of the knee joint is only guaranteed with biomechanically correct alignment.

8.1.1 Settings with the "X-Soft" adjustment software

8.1.1.1 Introduction

The "X-Soft" adjustment software makes it possible to optimise the product settings for a patient. The adjustment
software provides step-by-step guidance through the adjustment process. After the settings are configured, the
data for them can be saved and printed for documentation. If needed, these data can be loaded and fed into the
product.
You can consult the integrated online help of the adjustment software for further information.
INFORMATION
The 4X1 X-Soft adjustment software , version 1.4 or higher is required for correct alignment. If X-Soft is on
hand in a version of 1.0 or higher, it can be updated.
Updating the X-Soft adjustment software
1) Make sure you are connected to the internet and go to the "http://ottobock.com/en/datastation" website.
→ The webpage for the "Data Station" software will open.
2) Search for the description of the software that is to be updated under "All".
3) Search for the applicable version.
4) Click "Download" in the column on the right in order to download the update.
5) Extract the "ZIP file" and execute it.

8.1.1.2 Data transfer between the product and the PC

Product settings using the adjustment software can only be made via Bluetooth data transfer. For this, a Bluetooth
wireless connection must be established between the product and the PC with the help of the 60X5 BionicLink
PC. To install the 60X5 BionicLink PC for the first time, follow the procedure described in the instructions for use
supplied with this adapter.

8.1.1.3 Preparing the product to connect to the adjustment software

If the product does not emit any signals when querying the charge level (see Page 22), the battery is drained or the
product is switched off.
Switching on the product
1) Connect the power supply with battery charger to the wall socket.
2) Connect the battery charger to the product.
3) Wait for feedback signals.
4) Disconnect the battery charger from the product.
→ After feedback signals are emitted (self test), the product is switched on.
Switching on Bluetooth
Upon delivery the Bluetooth function of the prosthesis is switched on.
The Bluetooth function can be switched off using the cockpit app or the remote control. When the Bluetooth func­
tion is switched off, it is only turned on for 2 minutes after connecting/disconnecting the battery charger and is then
turned off again automatically. When a connection with the PC is active (the
function is not switched off automatically.
Genium X3 3B5-2/3B5-2=ST
Preparation for use
symbol is lit up), the Bluetooth