Speed Parameters - Otto Bock 490E75 Series Service Manual

Table of Contents

Advertisement

6.3.2

Speed parameters

Parameter
Setting range
Acceleration
Single steps
from 0 to 100
Deceleration
Single steps
from 0 to 100
Turn acceleration Single steps
from 0 to 100
Turn deceleration Single steps
from 0 to 100
Forward speed
Single steps
from 0 to 100
Reverse speed
Single steps
from 0 to 100
Turning speed
Single steps
from 0 to 100
Power
Single steps
from 0 to 100%
490E75=0_C
Description
Setting of forward and backward acceleration of the wheelchair.
A high value enables fast acceleration. This programmed acceleration value
appears when a speed setting of 5 has been selected. Its value for other set-
tings depends on the value of the parameter "Minimum acceleration".
Setting of the forward and backward deceleration (or brake) of the wheelchair.
A high value enables rapid deceleration. This programmed deceleration va-
lue appears when a speed setting of 5 has been selected. Its value for other
settings depends on the value of the parameter "Minimum deceleration".
Setting of turn acceleration of the wheelchair.
A high value enables fast acceleration. This programmed acceleration value
appears when a speed setting of 5 has been selected. Its value for other
settings depends on the value of the parameter "Minimum turn acceleration".
Setting of the turn deceleration (or brake) of the wheelchair.
A high value enables rapid deceleration. This programmed deceleration va-
lue appears when a speed setting of 5 has been selected. Its value for other
settings depends on the value of the parameter "Minimum turn deceleration".
Setting of the minimum and maximum forward speeds of the wheelchair.
A high value enables a high speed. The minimum value appears if the speed
setting 1 has been selected and the maximum value appears when the speed
setting 5 has been selected.
Setting of the minimum and maximum reverse speeds of the wheelchair.
A high value enables a high speed. The minimum value appears if the speed
setting 1 has been selected and the maximum value appears when the speed
setting 5 has been selected.
Setting of the minimum and maximum turning speeds of the wheelchair.
A high value enables a high speed. If "Drive profile 0" has been selected the
minimum value appears for speed setting 1 and the maximum value appears
when speed setting 5 has been selected.
Setting of wheelchair power level. Power is understood to be the ability, e.g.
to negotiate hills or overcome obstacles. If the power is set to 100% the
wheelchair can be operated at its full power. Values below 100% will result
in reduced power.
Power settings are usually made to avoid damage to corridors and furniture
when the wheelchair is being used indoors. The driving profiles can be set
independently of each other; i.e. different profiles can be defined for inside
and outside areas.
Programming example:
Maximum current = 70 A
Power (profile 1) = 100%
Power (profile 2) = 50% (i.e., the VR2 control has a level of 70 A in profile
1, and a level of 50% of 70 A = 35 A in profile 2).
VR2 wheelchair control