Manual, F/T Sensor, Ethernet Axia
Document #9620-05-C-Ethernet Axia-02
5.2 LED Normal Operation
5.2.1 Sensor Status LED
One LED signals the health status of the sensor as follows:
LED Color
(flash at 1 Hz
(flash at 10 Hz
Note:
1.
5.2.2 Diag LED
One LED signals the diagnostic status of the Ethernet Axia sensor interface as follows:
LED Color
Green Blinking Pre-operational
5.2.3 Ethernet Link/Activity LED
One LED signals link/activity on the communications port as follows:
LED Color
Pinnacle Park • 1031 Goodworth Drive • Apex, NC 27539 • Tel:+1 919.772.0115 • Fax:+1 919.772.8259 •
State
Off
No power
Green
Normal operation
Sensing range
1
Amber
exceeded
Red
Calibration error
speed)
Red
Communication error
speed)
Red
Status code error
(solid)
Amber is when both green and red LEDs are on.
State
Green
Operational
Red
Error
State
No power or no link
Off
activity
Green
Link activity
Table 5.2—Sensor Status LED
Description
Electricity is not supplied to the sensor.
The sensor's electronics are functioning and
communicating.
Indicates that an F/T axis is out of range.
Reduce the applied load or use a larger
calibration if available.
Sensor is not referencing a calibration range
or has a checksum error.
The sensor is not able to communicate data
over the communication protocol.
For more information on the error set,
refer to
Table
Table 5.3—Diag LED
Description
Defined by the communication/protocol
standard.
No errors are found.
Indicates an error reported by the internal
electronic components.
Also, after a UART error, the LED stays red for
five seconds.
Table 5.4—L/A LED
Description
Link/activity is not detected.
This light stays green
for 5 seconds after any link activity.
C-26
5.7.
www.ati-ia.com