Storz IMAGE1 S CONNECT Instructions For Use Manual page 58

Hide thumbs Also See for IMAGE1 S CONNECT:
Table of Contents

Advertisement

9.7 Table 5 – Recommended separation distances between
portable and mobile HF communications devices and
the product
The product is intended for use in an electromagnetic environment in which HF disturbances
are controlled. The customer or user of the product can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile HF
communications equipment (transmitters) and the product as recommended below, according
to the output energy of the communications equipment.
Rated maximum
output power of the
transmitter [W]
For transmitters whose maximum rated power is not listed in the table above, the recom-
mended separation distance d in meters (m) can be estimated using the equation from the re-
spective column, whereby P is the maximum rated power of the transmitter in watts (W) ac-
cording to the transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for higher frequency range applies.
Note: These guidelines may not apply in all situations. The propagation of electromagnetic
waves is affected by absorptions and reflections of buildings, objects, and people.
The product was tested for compatibility with HF surgical devices in accordance with IEC
60601-2-2 Appendix BB.
Instructions for use • IMAGE1 S • 705LZA_EN_V2.1_05-2022_IFU_FDA
Separation distance d [m] according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2 √P
0.01
0.1
1
10
100
Electromagnetic compatibility
80 MHz to 800 MHz
d = 1.2 √P
0.12
0.38
1.2
3.8
12
800 MHz to 2.7 GHz
d = 2.3 √P
0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23
58

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Image1 s connect iiImage1 s h3-linkImage1 sx-linkImage1 s d-linkImage1 s 4u-linkImage1 s d3-link extension cable ... Show all

Table of Contents