5.6 Performing the white balance

The white balance must be performed again after replacing the endoscope.
Failure to perform proper white balance of system may result in color distortion.
1.
Ensure that the imaging instrument is connected and a live image appears on the
monitor.
2.
Connect the light source to the imaging instrument via the light cable.
3.
Point the tip of the imaging instrument at a white and clean surface, e.g., white gauze.
4.
Make sure that the light source emits enough light to sufficiently illuminate the surface.
ð The monitor displays a completely white and well illuminated live image.
5.
Press the White Balance button on the IMAGE 1 CONNECT or the button in the Live
menu.
ð The white balance is performed.
6.
Check the monitor to see whether the white balance was successful.
7.
Repeat the white balance if required.
Instructions for use • IMAGE1 S • 705LZA_EN_V2.1_05-2022_IFU_FDA