Storz IMAGE1 S CONNECT Instructions For Use Manual page 54

Hide thumbs Also See for IMAGE1 S CONNECT:
Table of Contents

Advertisement

Type
LINK cable (IML
cable)
BNC to BNC (12G-
SDI cable)
DVI
4-pin Mini DIN
Power cord
Ethernet - CAT6A
9.3 Table 1 – Compliance level for immunity tests
Guidelines and manufacturer's declaration – electromagnetic immunity
The product is intended for use in the electromagnetic environment specified below. The user
of the product should make sure that it is used in such an environment.
Interference im-
munity tests
Electrostatic dis-
charge (ESD) acc.
to IEC 61000-4-2
Electrical fast
transients/bursts
acc. to IEC
61000-4-4
Surges acc. to
IEC 61000-4-5
Voltage dips,
short interrup-
tions, and voltage
variations acc. to
IEC 61000-4-11
Instructions for use • IMAGE1 S • 705LZA_EN_V2.1_05-2022_IFU_FDA
System cables and maximum lengths used for EMC compliance.
Shield
Längth [m]
Yes
2
Yes
10
Yes
3
Yes
1
No
2
Yes
2
EN/IEC 60601 test
level
± 8 kV contact dis-
charge
± 15 kV air discharge
± 2 kV for power lines
± 1 kV for input and
output lines
100 kHz repetition
± 1 kV voltage outer
conductor – outer con-
ductor
± 2 kV voltage outer
conductor – ground
Voltage dip:
Dip to 0% for 1 cycle
at 0° phase angle
Dip to 70% for 25/30
cycles at 0° phase an-
gle
Dropout to 0% for 0.5
cycles @ 0°, 45°, 90°,
135°, 180°, 225°, 270°,
and 315° phase angles
Voltage interruption:
Electromagnetic compatibility
Ferrite
Use
No
Interconnection of intermodule
LINK signals
No
Connection of 3G/12G-SDI video
signals
Yes
Connection of DVI signals
No
Connection of SCB signals
No
Connection of device to AC
Mains
No
Connection of HIVE signals
Compliance level
± 8 kV contact dis-
charge
± 15 kV air discharge
± 2 kV for power lines
± 1 kV for input and
output lines
100 kHz repetition
± 1 kV voltage outer
conductor – outer con-
ductor
± 2 kV voltage outer
conductor – ground
Voltage dip:
Dip to 0% for 1 cycle
at 0° phase angle
Dip to 70% for 25/30
cycles at 0° phase an-
gle
Dropout to 0% for 0.5
cycles @ 0°, 45°, 90°,
135°, 180°, 225°, 270°,
and 315° phase angles
Voltage interruption:
Electromagnetic envi-
ronment – guidelines
Floors should be made
of wood, concrete, or
covered with ceramic
tiles. If floors are cov-
ered with synthetic ma-
terial, the relative hu-
midity must be at least
30%.
The power supply qual-
ity should be that of a
typical commercial or
hospital environment.
The power supply qual-
ity should be that of a
typical commercial or
hospital environment.
The power supply qual-
ity should be that of a
typical commercial or
hospital environment. If
the user of the product
requires continued op-
eration in the event of
interruptions to the
power supply network,
it is recommended that
the product be oper-
54