Test fre-
quency
MHz
1720
1845
1970
2450
5240
5500
5785
9.5 Table 3 – Test levels for radiated and conducted
immunity tests
Guidance and manufacturer's declaration – Electromagnetic immunity
The product is intended for use in the electromagnetic environment specified below. The user
should make sure that it is used in such an environment.
Immunity tests
Conducted RF distur-
bances acc. to IEC
61000-4-6
Radiated RF distur-
bances acc. to IEC
61000-4-3
Instructions for use • IMAGE1 S • 705LZA_EN_V2.1_05-2022_IFU_FDA
Frequency
Radio service Modulation
band
MHz
1700 – 1990
GSM 1800,
CDMA 1900,
GSM 1900,
DECT,
LTE band 1, 3,
4, 25,
UMTS
2400 – 2570
Bluetooth,
WLAN 802.11
b/g/n,
RFID 2450,
LTE Band 7
5100 – 5800
WLAN 802.11
a/n
EN/IEC 60601 test
level
3 V
rms
150 kHz to 80 MHz
6 V
in ISM ands be-
eff
tween 0.15 MHz and
80 MHz
3 V/m
80 MHz to 2.7 GHz
Electromagnetic compatibility
Immunity
test level
V/m
Pulse modula-
28
tion
217 Hz
Pulse modula-
28
tion
217 Hz
Pulse modula-
9
tion
217 Hz
Compliance
Electromagnetic envi-
level
ronment – Guidance
3 V
Portable and mobile RF
rms
communications equip-
150 kHz to
ment should be used no
80 MHz
closer to any part of the
6 V
in ISM
product, including cables,
eff
ands between
than the recommended
0.15 MHz and
safety distance calculated
80 MHz
from the equation applica-
ble to the frequency of the
3 V/m
transmitter.
80 MHz to
Recommended safety dis-
2.7 GHz
tances:
d = 1.2 √P
Where P is the rated
power of the transmitter in
watts [W] according to the
information provided by
the transmitter manufac-
turer and d is the recom-
mended safety distance in
meters [m].
Compliance
level
V/m
28
28
9
56