9 Electromagnetic compatibility

9.1 General notes on the operating environment

The product is suitable for use in professional healthcare settings. Professional healthcare
facilities include physician offices, dental offices, limited care facilities, freestanding surgical
centers, freestanding birth centers, multiple treatment facilities, hospitals (emergency rooms,
patient rooms, intensive care, surgical rooms, outside the HF-shielded room of an ME system
for MRT).
The emission characteristics of this product make it suitable for use in industrial areas as
well as in hospitals (CISPR 11 Class A) and other professional healthcare environments.
If it is used in a residential environment (for which CISPR 11 Class B is normally
required), the product may not offer sufficient protection for radio transmission operation.
The user might need to take mitigation measures, such as relocating or re-orienting the
product.
WARNING
MR unsafe!
This product is MR unsafe.
Keep the product away from magnetic resonance Imaging scanner room
WARNING
Electromagnetic interferences! Malfunction!
Use of this equipment adjacent to or stacked with other equipment could result in improper
operation.
Avoid this situation.
If such use is necessary: Ensure that this equipment and the other equipment are
operating normally.
WARNING
Reduced immunity or increased emissions! Malfunction!
Use of the product with accessories, transducers and cables other than those specified in this
manual may result in increased emissions or decreased immunity.
Only use the accessories specified in the manual.

9.2 Accessories and cables

WARNING
Degradation of performance! Malfunction!
Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) could result in degradation of the performance of the product.
Do not use portable communications equipment closer than 30 cm (12 inches) to any
part of the product, including cables.
Instructions for use • IMAGE1 S • 705LZA_EN_V2.1_05-2022_IFU_FDA
Electromagnetic compatibility
53