5.5 Putting the product into operation .................. 24 5.5.1 Switching the product on and off ............... 25 5.5.2 Performing the white balance ................25 5.5.3 Performing the function test ................26 Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
Page 4
9.6 Table 4 – Emission class and group ................44 9.7 Table 5 – Recommended separation distances between portable and mobile HF communications devices and the product ..............45 10 Errors and messages ........................ 46 10.1 Troubleshooting ....................... 46 Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
Page 5
Table of contents 11 Subsidiaries ..........................47 Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
ð Interim result of an action Step 2 ð Result of a completed action Actions in safety notes or in the case of a single step: Step 1 Lists Numbered list – Unnumbered list, 1st level Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
Designates a possible imminent risk. If this is not avoided, it could lead to minor injuries. NOTICE NOTICE Designates a possibly harmful situation. If this is not avoided, the products could be damaged. Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
2 Normal use 2.1 Intended use The TELECAM C3 is a camera control unit for use with camera heads or videoendoscopes and is used for visualizing and documenting endoscopic and microscopic procedures. The TELECAM C3 has no direct contact with the human body.
All electrical installations in the operating room in which the product is connected and used must meet applicable IEC standards. Do not touch the output jacks of the product and the patient at the same time during use Do not open the product. Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
Safety and warning Use either the power cord supplied by KARL STORZ or a power cord which has the same properties and which bears a national mark of conformity. Make sure that the plug is completely clean and dry. To ensure reliable protective earth grounding, connect the product to a properly installed socket that is approved for use in the operation room.
4 Product description 4.1 Product overview The TELECAM C3 is a camera control unit for endoscopic image display and documentation. An intuitive user interface ensures simple and quick access as well as rapid understanding of menu navigation. Click here for a detailed description of the functions see chapter User interface [p. 28].
TH110 IMAGE1 S HX-P TH111 IMAGE1 S D1 TH115 IMAGE1 S D1 TH116 Compatible camera heads for C-Line connection Article Order no. H1 Camera head TH130 Compatible videoendoscopes for X-Line connection Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
Page 13
Standard colonoscope 3.8/1400 NTSC, set Standard colonoscope 3.8/1600 PAL 13925PKS, 13925NKS Standard colonoscope 3.8/1600 NTSC Medium gastroscope 7.8/2.8/1100 PAL 13823PKS, 13823NKS Medium gastroscope 7.8/2.8/1100 NTSC Slim colonoscope 11.2/3.4/1330 PAL 13926PKS/NKS Slim colonoscope 11.2/3.4/1330 NTSC Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
Page 14
C-MAC S Imager 8403XSI C-MAC S Pediatric Imager 8403XSP C-MAC Video laryngoscope MAC #0 8403EXC C-MAC Video laryngoscope MAC #2 8403KXC C-MAC Video laryngoscope MAC #3 8403AX C-MAC Video laryngoscope MAC #3 8403AXC Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
FIVE S 5.3x65, sterile for single use 0915612-01 Video-uretero-renoscope FLEX-X 091271-01 Video-uretero-renoscope FLEX-X 091279-01 4.3 Technical data TELECAM C3 TC100 Power supply (AC) 100–240 V Operating frequency 50/60 Hz Power input 100–135 VA Electrical protection class Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
4.4.1 Symbols on the packaging Symbol Meaning Manufacturer Date of manufacture Medical device Article no. Serial number Number of products in the product packaging Unique Device Identifier Consult the printed or electronic instructions for use Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
The color may differ on the product. The symbol is black/white on the packaging label. Ready/standby button Applied part type BF White balance Applied part type CF defibrillation protection Potential equalization connector Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
Exits the Quick menu. Alternatively, exit the menu using the ESC button on the keyboard. The Close button appears only if a keyboard is used or the product is op- erated via the camera head buttons. Still Captures a still image. Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
Page 19
Selects the image refresh rate between 50 and 60 Hz for the DVI output. The product has to be restarted after the frequency is changed. Date and time Sets the date, time, and display format. Patient Management Manages the settings for handling patient data. Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
Page 20
Calls up the storage space currently used as well as the free storage space. Internal Storage Calls up the storage space currently used as well as the free storage space. Transfers stored data to a USB storage device. Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
5.4.1 Connecting the potential equalization The product’s ground line should be installed by a qualified electrician. Connect the potential equalization cable to the plug connection for potential equalization on the device. Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
5.4.2 Connecting the monitor ü The connected monitor supports a resolution of 1920 x 1080 pixels. Connect the monitor to the TELECAM C3 via one of the DVI outputs on the rear of the device. Connect the primary work screen to DVI1.
Before connecting the devices to the power supply, check that the values on the power sockets match the values on the device's type plate. Insert the device plug of the power cord into the TELECAM C3 power socket. Connect the power plug of the TELECAM C3 to a power socket.
ð The light source emits enough light to sufficiently illuminate the white surface. The monitor displays a completely white and well illuminated live image. Press the White Balance button on the front of the TELECAM C3 or the White Balance button in the live menu to perform the white balance.
5.5.3 Performing the function test ü Perform the function test before each use. Check the TELECAM C3 and the camera head or videoendoscope for external damage. Ensure that the camera head cable or videoendoscope cable is not damaged or kinked.
Camera head button functions Button Function Scroll down Scrolls down the menus or options Scroll up Scrolls up the menus or options Menu button Activates a highlighted option Selects the Quick menu Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
Press the corresponding symbol for the desired function. Activate the desired function. 6.3.2 Quick menu configuration The Quick menu configuration window is divided into three columns. The functions can be moved, replaced, added or removed. Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
6.5 Setup menu The general settings can be changed in the Setup menu. 6.5.1 Accessing the Setup menu Press the Function Space button. Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
Calls up all the necessary information on the system and the connected camera heads or videoendoscopes. Calls up information about a connected com- patible camera head or a connected compati- ble videoendoscope: – Item number – Item name Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
Select the connected product from Connected Devices in the Information panel. ð The camera head button assignment is shown true to original. Press the Settings button. Change the button assignment as required. Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
All safety-relevant actions are recorded and saved by the system. KARL STORZ assumes no legal liability or responsibility for the loss, accuracy, damage, or disclosure of patient data that has been managed using the TELECAM C3.
Page 33
Anonymous recording of media files Images or video recordings are stored anony- mously. 6.8.1.3 Editing the storage limit Call up the Storage Limit sub-menu. Specify the desired setting for data storage. Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
Specify the desired setting for data export. Option Function USB (default setting) Exporting data to a USB storage device is al- lowed. No Export Exporting data to a USB storage device is not allowed. Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
The following roles can be selected: – Administrator: Full access to all functions and settings – Operating room personnel: Access to patient data. No access to patient data security or storage medium Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
Technician: Access to device settings, e.g. updates. No access to patient data or other sensitive areas. All safety-relevant actions are recorded and saved by the system. KARL STORZ assumes no legal liability or responsibility for the loss, accuracy, damage, or disclosure of patient data that has been managed using TELECAM C3.
Document the results of the safety inspection. 7.4 Repairing the product Repair work may only be performed by KARL STORZ or by a company authorized by KARL STORZ. The interventions described in this instruction manual are exempt from this rule. Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
Maintenance, servicing, repairs, and disposal Please contact your local KARL STORZ subsidiary or authorized dealer (see the list of subsidiaries). Contaminated devices may not be shipped. To prevent contact infections and airborne infections, products must first be decontaminated. KARL STORZ reserves the right to send back contaminated products.
USB silicone keyboard with touchpad, IT 20040241IT USB silicone keyboard with touchpad, PT 20040241PT USB silicone keyboard with touchpad, RU 20040241RU USB silicone keyboard with touchpad, SE 20040241SE USB silicone keyboard with touchpad, US 20040241US Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
Do not use portable communications equipment closer than 30 cm (12 inches) to any part of the product, including cables. Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
Page 41
@ 0°, 45°, 90°, it is recommended that 135°, 180°, 225°, 270°, 135°, 180°, 225°, 270°, the product be oper- and 315° phase angles and 315° phase angles Voltage interruption: Voltage interruption: Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
Page 42
704 – 787 LTE band 13 Pulse modula- and 17 tion 217 Hz 800 – 960 GSM 800/900, Pulse modula- TETRA 800, tion iDEN 820, 18 Hz CDMA 850, LTE band 5 Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
Page 43
Where P is the rated power of the transmitter in watts [W] according to the information provided by the transmitter manufac- turer and d is the recom- mended safety distance in meters [m]. Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
Page 44
IEC 61000-3-2 public low voltage power supply network that supplies buildings Voltage fluctuations/flicker emis- Complies used for domestic purposes. sions acc. to IEC 61000-3-3 Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
Page 45
Note: These guidelines may not apply in all situations. The propagation of electromagnetic waves is affected by absorptions and reflections of buildings, objects, and people. The product was tested for compatibility with HF surgical devices in accordance with IEC 60601-2-2 Appendix BB. Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...
Loss of image for longer than Defibrillator discharged. Switch the device off 4 seconds and back on again. Update failed Restart the device Repeat the procedure Contact Service Incompatible camera head Replace the camera head Instructions for use • TELECAM C3 TC100 • LZA600_EN_V2.0_02-2022_IFU_CE-MDR...