Product-Specific - KaVo Primus 1058 Life TM Instructions For Use Manual

Hide thumbs Also See for KaVo Primus 1058 Life TM:
Table of Contents

Advertisement

Instructions for use Primus 1058 Life
2 Safety | 2.2 Purpose – Proper use
The services encompass all the test tasks required in accordance with § 6 of the
medical devices operator ordinance (Medizinprodukte-Betreiberverordnung, MP-
BetreibV).
Note
The product must be cleaned and serviced according to instructions if it is not
to be used for an extended period of time.
Note
The MULTIflex couplings, the current K/KL motors, and the ultrasonic scaler
hoses of KaVo are equipped as standard with a protective device to prevent
treatment water from being drawn back into the treatment centre via the
handpieces. If products from other manufacturers are used at the standard-
ised interfaces, it must be ensured that they are equipped with an appropri-
ate protective device! If this is not the case, they may not be used!
Information about electromagnetic compatibility
Note
Based on IEC 60601-1-2 (DIN EN 60601-1-2) concerning the electromag-
netic compatibility of electrical medical devices, we must draw your attention
to the following points:
• Medical electrical devices are subject to special precautions concerning the
electromagnetic compatibility and must be installed and operated in accord-
ance with the KaVo assembly instructions.
• High-frequency communications devices may interfere with electrical med-
ical devices.
See also:
2 10 Data on electromagnetic compatibility according to EN 60601-1-2, Page
119
Note
KaVo cannot guarantee the compliance of accessories, cables, and other
components not supplied by KaVo with the EMC requirements of IEC
60601-1-2 (DIN EN 60601-1-2).
Disposal
Note
Any waste which is generated must be recycled or disposed of in strict com-
pliance with all applicable national regulations in a manner which is safe both
for people and the environment.
If you have any questions regarding proper disposal of the KaVo product,
please contact the KaVo branch.
Disposal of electronic and electrical devices

2.2.2 Product-specific

Designated use and target group
KaVoPrimus 1058 Life is designed for dental treatment of children and adults.
12 / 126

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Kavo primus 1058 life c

Table of Contents