Microlife BP A7 Touch BT Manual page 11

Hide thumbs Also See for BP A7 Touch BT:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 3
Electromagnetic Compatibility Information
This device is compliant with EN60601-1-2: 2015 Electromagnetic
Disturbances standard.
This device is not certified to be used in vicinity of High Frequency
(HF) medical equipment.
Do not use this device close to strong electromagnetic fields and
portable radio frequency communication devices (for example
microwave oven and mobile devices). Keep a minimum distance
of 0.3 m from such devices when using this device.
Device care
Clean the device only with a soft, dry cloth.
Cleaning the cuff
Carefully remove spots on the cuff with a damp cloth and soap-
suds.
m
WARNING: Under no circumstances may you wash the
inner bladder!
Accuracy test
We recommend this device is tested for accuracy every 2 years or
after mechanical impact (e.g. being dropped). Please contact your
local Microlife-Service to arrange the test (see foreword).
Disposal
Batteries and electronic devices must be disposed of in
accordance with the locally applicable regulations, not with
domestic waste.
12. Guarantee
This device is covered by a 5 year guarantee from the date of
purchase. During this guarantee period, at our discretion, Microlife
will repair or replace the defective product free of charge.
Opening or altering the device invalidates the guarantee.
The following items are excluded from the guarantee:
 Transport costs and risks of transport.
 Damage caused by incorrect application or non-compliance
with the instructions for use.
 Damage caused by leaking batteries.
 Damage caused by accident or misuse.
 Packaging/storage material and instructions for use.
 Regular checks and maintenance (calibration).
 Accessories and wearing parts: Batteries, power adapter
(optional).
Microlife BP A7 Touch BT
The cuff is covered by a functional guarantee (bladder tightness)
for 2 years.
Should guarantee service be required, please contact the dealer
from where the product was purchased, or your local Microlife
service. You may contact your local Microlife service through our
website:
www.microlife.com/support
Compensation is limited to the value of the product. The guarantee
will be granted if the complete product is returned with the original
invoice. Repair or replacement within guarantee does not prolong
or renew the guarantee period. The legal claims and rights of
consumers are not limited by this guarantee.
13.Technical specifications
Operating condi-
10 - 40 °C / 50 - 104 °F
tions:
15 - 90 % relative maximum humidity
Storage conditions: -20 - +55 °C / -4 - +131 °F
15 - 90 % relative maximum humidity
Weight:
312 g (including batteries)
Dimensions:
160 x 82 x 35 mm
Cuff size:
from 17 - 52 cm according to the cuff sizes
(see «Selecting the correct cuff»)
Measuring proce-
oscillometric, corresponding to Korotkoff
dure:
method: Phase I systolic, Phase V
diastolic
Measurement range: 20 - 280 mmHg – blood pressure
40 - 200 beats per minute – pulse
Cuff pressure display
range:
0 - 299 mmHg
Resolution:
1 mmHg
Static accuracy:
within ± 3 mmHg
Pulse accuracy:
± 5 % of the readout value
Communication:
Bluetooth
Compatibility:
iOS: iOS 10.0 or newer
Android: Android : Android OS 5 or newer
Voltage source:
4 x 1.5 V alkaline batteries; size AAA
Mains adapter DC 6V, 600 mA (optional)
Battery lifetime:
approx. 400 measurements
(using new batteries)
®
4.0
9
EN

Advertisement

Table of Contents
loading

Table of Contents