lower-
limited
highly-
limited
start
voltage
start
time
end time Test time of end voltage, range:
count
Save To
File
Copyright © Itech Electronic Co., Ltd.
Power Supply Function
V4: Voltage in interval 4
V5: Voltage in interval 5
V6: Voltage in interval 6
Range of lower limited operating ca-
pability, with the following settings:
V1: Voltage in interval 1
V2: Voltage in interval 2
V3: Voltage in interval 3
V4: Voltage in interval 4
V5: Voltage in interval 5
V6: Voltage in interval 6
Range of highly limited operating ca-
pability, with the following settings:
V1: Voltage in interval 1
V2: Voltage in interval 2
V3: Voltage in interval 3
V4: Voltage in interval 4
V5: Voltage in interval 5
Set the initial output voltage of the in-
strument before the LV123 regulation
waveform test. At the same time, this
value is also the end voltage after the
regulatory waveform test is
completed.
Test time of starting voltage, range:
0~999.999S.
0~999.999S.
Set the number of test repetitions,
ranging from 1 to 65535. The actual
number of test executions = count *
3.
Set the save address of the waveform
file, range: 1~10.
93