Presets
Use this section to change the preset values within the High Voltage, Magnification, Stage and Pressure tabs.
• High Voltage – values must be entered in kilovolts (0.2 kV = 200 V) and span from 350 V to 30 kV.
• Magnification – can be changed to hold frequently used magnifications from 20× to 1 000 000×. Values that are
in the list but cannot be applied to the actual SEM conditions are not shown in the toolbar Magnification list box.
Note
Alternatively, magnification could be shown as the Horizontal Field Width (see the Preferences... dialog / General section /
Image dimensions control item).
• Stage – Tilt and Rotation presets can be set in Absolute or Relative values.
• Pressure – list can be changed to hold specific values frequently used in the Low Vacuum / ESEM mode within
the range of 10 to 4 000 Pa.
Presets preferences
FIGURE 3-19
Click the Add button, add new value and confirm the entry; the new value is sorted into the list. Click the Delete
button to delete selected value. The Default button on each tab sets the factory presets values.
C O N F I D E N T I A L – limited rights
Revision A
Feb 2018
Software control: Preferences dialog
User Manual
3-39