Emergency off
To switch off the electrical power completely in case of emergency, follow this quick and safe procedure:
1. Push the red EMERGENCY OFF (EMO) button (option – see the Apreo User safety manual).
If the button is not installed proceed as follows:
2. Switch off the breaker switch labeled Main switch S1 at the cabinet back, which is placed at the very left side of
the row of three, or disconnect the power cord labeled Main input 230 VAC.
Mains switch board
FIGURE 2-3:
If this is not easily accessible:
3. Turn off the mains wall switch (if present), and / or disconnect the mains plug from the mains socket.
W A R N I N G !
The microscope console water chiller (option – separate device in the microscope vicinity) is powered separately via its
individual power cord. Hazardous voltages may be present in this equipment even when the microscope power plug is
disconnected!