Stage movement limits
The motorized movements of the stage can be operated under software control for more advanced location
mapping. This includes Shift, Get, Track and Stage module functionality. A live image can be repositioned either by
the stage movement or by the Beam Shift.
Note
When moving the stage or tilting the specimen, the magnification may need to be reduced so as to not lose the feature of
interest from the screen.
Stage movement schema
FIGURE 2-7:
Stage features and limits
Table 2-3:
Item
X
Y
Z
R
T
Eucentric distance
Maximum sample weight
C a u t i o n !
The positive Z value direction depends on the Link Z to FWD status (see Chapter 5).
C a u t i o n !
If the maximum sample size is near the limit, stage tilt can be limited. Beware of hitting the objective pole piece!
2-16
User Manual
System control: Equipment
110 × 110 mm
-55 to +55 mm
-55 to +55 mm
66 mm
360°
-15° to +90°
10 mm
No restriction: 500 g
No Tilt: 2 000 g
X and Y movements only: 5 000 g
Note
continuous
maximally
including stub holder
C O N F I D E N T I A L – limited rights
Feb 2018
Revision A