Operating Procedures: Stage Control
Navigation Alignment
This procedure, in comparison to the Navigation
montage or Nav-Cam (option), enables using any
loaded or paused Navigation image that is
calibrated according to the live Reference image.
Follow the instructions given within the process
(6 dialogs) and calibrate 4 image points. If 2 points
are sufficient for the desired purpose, click on the
Compute 2pt alignment button to finish the process
after setting 2 alignment points.
When any point setting does not satisfy system
requirements, a Warning message informs the user.