Diagnostics And Troubleshooting; General Troubleshooting - Endress+Hauser Proline Promass E TB2 HART Operating Instructions Manual

Coriolis mass flow measuring system
Table of Contents

Advertisement

Proline Promass E TB2 HART
Endress+Hauser
12

Diagnostics and troubleshooting

12.1

General troubleshooting

Problem
Local display dark and no signal
output at current output (0 mA)
Local display is dark, but signal
output is within the valid current
range (3.6 to 22 mA)
Signal output outside the valid
current range
(< 3.6 mA or > 22 mA)
Device shows correct value on local
display, but signal output is
incorrect, though in the valid
current range
Device measures incorrectly.
Text on measured value display and
in navigation view appears in a
foreign language and cannot be
understood
No connection via HART protocol
No connection via CDI interface
Possible cause
Supply voltage does not match that
specified on the nameplate.
Supply voltage has incorrect polarity.
No contact between connecting cables
and terminals.
Terminals are not plugged into the I/O
electronics module correctly.
I/O electronics module is defective.
Display is set too bright or too dark.
Spiral cable of the display module is not
plugged in correctly.
Display module is defective.
Main electronics module is defective.
I/O electronics module is defective.
Configuration error
Configuration error or device is operated
outside the application.
Incorrect operating language is
configured.
Missing or incorrectly installed
communication resistor.
Commubox
• Connected incorrectly
• Configured incorrectly
• Drivers not installed correctly
• USB or COM interface on computer
configured incorrectly
Incorrect configuration of USB interface
on computer or PC or driver not installed
correctly.
Diagnostics and troubleshooting
Remedy
Apply the correct supply voltage
(® ä 26).
Reverse polarity of supply voltage.
Check the contacting of the cables and
correct if necessary.
Check terminals.
Order spare part (® ä 91).
• Set the display brighter by
simultaneously pressing O + F.
• Set the display darker by
simultaneously pressing S + F.
Insert the plug correctly into the main
electronics module and display module.
Order spare part (® ä 91).
Order spare part (® ä 91).
Order spare part (® ä 91).
Check and correct parameter
configuration.
1. Check and correct parameter
configuration.
2. Observe limit values specified in the
"Technical Data".
1. Press 2 s S + O ("home position").
2. Press F.
3. Set the desired language in the
Language parameter.
Install the communication resistor (250
W) correctly. Observe the maximum load
(® ä 27).
Observe the documentation for the
Commubox.
• FXA 191 HART: Document
"Technical Information"
TI237F/00/xx.
• FXA 195 HART: Document
"Technical Information"
TI404F/00/xx.
Observe the documentation for the
Commubox.
FXA 291 HART: Document
"Technical Information" TI405C/
07/xx.
81