Commissioning
64
10.4.5

Configuring the low flow cut off

The Low flow cut off wizard guides you systematically through all parameters that have to be set
for configuring the low flow cut off.
Navigation path
"Setup" menu ® Low flow cut off
Sequence of the wizard
Off
å 16
"Low flow cut off" wizard in the "Setup" menu
Parameter overview with brief description
Parameters
Assign process variable
Select the process variable for
low flow cut off.
On value low flow cut off
Enter the on-value for the
low flow cut off.
Off value low flow cut off
Enter the off-value for the
low flow cut off.
Pressure shock suppression
Enter the time interval for
signal suppression (= active
pressure shock suppression).
Assign process
variable
Mass flow
Volume flow
On value
low flow
cutoff
Off value
low flow
cutoff
Pressure shock
suppression
End of wizard
Description
Selection/
User entry
• Off
• Mass flow
• Volume flow
• Corrected volume flow
Max. 15-digit, positive
floating-point number
0 to 100 %
0 to 100 s
Proline Promass E TB2 HART
Corrected volume
flow
A0013799-EN
Factory setting
Mass flow
For liquids: depends on
country and nominal
diameter
50 %
0 s
Endress+Hauser