While this action is in progress, the configuration cannot be edited via the local display and a
message on the processing status appears on the display.
10.7

Simulation

The Simulation submenu enables you to simulate, without a real flow situation, various process
variables in the process and the device alarm mode and to verify downstream signal chains (switching
valves or closed-control loops).
Navigation path
"Diagnostics" menu ® Simulation
Structure of the submenu
Simulation
Parameter overview with brief description
Parameters
Prerequisite
Assign simulation
-
process variable
Value process variable One of the following
options is selected in
the Assign process
variable parameter:
• Mass flow
• Volume flow
• Corrected volume
flow
• Density
• Reference density
• Temperature
Simulation current
-
output 1-2
Value current output
The On option is
1-2
selected in the
Simulation current
output 1-2 parameter
®
Assign simulation process
variable
Value process variable
Simulation current output 1
Value current output 1
Simulation current output 2
Value current output 2
Simulation device alarm
Description
Select a process
variable for the
simulation process that
is activated.
Enter the simulation
value for the selected
process variable.
Switch simulation of
the current output on
and off.
Enter the current value
for simulation.
Commissioning
Selection/
Factory setting
User entry
• Off
Off
• Mass flow
• Volume flow
• Corrected volume
flow
• Density
• Reference density
• Temperature
Depends on the
-
process variable
selected
• On
Off
• Off
3.6 to 22.5 mA
Current value
currently measured
73