Diagnostics and troubleshooting
86
Diagnostics event
Event text
S044
Sensor drift
S046
Sensor limit
F062
Sensor connection
F082
Data storage
F083
Memory content
1)
Event level can be changed: Chapter 12.5.1 "Adapting event level"
12.5.3

Diagnostics for the electronics

Diagnostics event
Event text
F242
Software incompatible 1. Check software.
F252
Modules incompatible 1. Check electronic modules.
F261
Electr. modules
F262
Module connect.
F270
Main electronic
F271
Main electronic
F272
Main electronic
F273
Main electronics
S274
Main electronic
F275
I/O module
F276
I/O module
F282
Data storage
F283
Memory content
F311
Electronics error
M311
Electronics error
1)
Event level can be changed: Chapter 12.5.1 "Adapting event level"
Remedial measures
Electrodynamic sensor signal deviation!
1. Check or change main electronics.
2. Change sensor.
The measuring tube oscillation frequency is
outside the permitted range!
1. Inspect sensor.
2. Check process condition.
1. Change main electronic module.
2. Change sensor.
1. Change main electronic module.
2. Change sensor.
1. Restart device.
2. Restore S-Dat data.
3. Change sensor.
Remedial measures
2. Flash or change main electronics module.
2. Change I/O or main electronics module.
1. Restart device.
2. Check electronic modules.
3. Change I/O Modul or main electronics.
1. Check module connections.
2. Change electronics modules or sensor.
Change main electronics module.
1. Restart device.
2. Change main electronic module.
1. Restart device.
2. Contact service.
Emergency operation via display!
Change main electronics.
Unstable measurement!
Change the main electronics.
Change I/O module.
1. Restart device.
2. Change I/O module.
1. Restart device.
2. Contact service.
1. Transfer data or reset device.
2. Contact service.
1. Transfer data or reset device.
2. Contact service.
Maintenance required!
1. Do not perform reset.
2. Contact service.
Proline Promass E TB2 HART
Event level from the
factory
1)
Warning
1)
Warning
Alarm
Alarm
Alarm
Event level from the
factory
Alarm
1)
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
1)
Warning
Alarm
Alarm
Alarm
Alarm
Alarm
Warning
Endress+Hauser