Mounting
20
Ensure the system pressure is sufficiently high to prevent cavitation and outgassing.
For this reason, the following mounting locations are recommended:
å 6
Downstream from pumps (no danger of vacuum)
å 7
At the lowest point in a vertical pipe
Thermal insulation
Some fluids require suitable measures to reduce the heat convection from the sensor to the
transmitter. A wide range of materials can be used to provide the required thermal insulation.
NOTICE
Electronics overheating on account of thermal insulation!
Observe maximum permitted insulation height of the transmitter neck so that the transmitter
head is completely free.
Heating
NOTICE
Electronics can overheat due to elevated ambient temperature!
Observe maximum permitted ambient temperature for the transmitter (® ä 19).
Depending on the fluid temperature, take the device orientation requirements into account
(® ä 18).
Proline Promass E TB2 HART
mm (in)
Endress+Hauser
A0013935
A0013936
A0013937