Proline Promass E TB2 HART
Protocol-specific data
Terminal assignment
Supply voltage
Power consumption
Power supply failure
Electrical connection
Load
Potential equalization
Terminals
Cable entries
Cable specification
Endress+Hauser
• For information on the device description files (® ä 49)
• For information on the HART device variables and measured values
• For information on the HART commands
18.5

Power supply

Order version
Output
8E2B**-**C********
Order version,
output
8E2B**-**C********
*
Supply voltage range without load. For information on the supply voltage with a load (® ä 27)
• Operation with current output 1: 65 to 660 mW
• Operation with current output 1 and 2: 130 to 1 320 mW
• Totalizers stop at the last value measured.
• Configuration is retained in the HistoROM.
• Error messages (incl. total operated hours) are stored.
(® ä 25)
(® ä 27)
No special measures for potential equalization are required.
Plug-in spring terminals for wire cross-sections 0.5 to 2.5 mm
• Cable gland: M20 × 1.5 with cable Æ 6 to 12 mm (0.24 to 0.47 in)
• Thread for cable entry: ½" NPT, G ½"
• Permitted temperature range: –40 °C (–40 °F) to³ 80 °C (176 °F); at least ambient temperature
+ 20 K
• A normal device cable suffices if only the analog signal is used.
• A shielded cable is recommended if using the HART protocol. Observe grounding concept of the
plant.
Terminal numbers
3 (+)
4 (-)
4-20 mA
U
*
S
(supply voltage)
18 to 30 V DC
2
(20 to 14 AWG)
Technical data
1 (+)
2 (-)
4-20 mA HART
I
max
(output current)
22.0 mA