Mounting

6.1

Mounting requirements

No special measures such as supports are necessary. External forces are absorbed by the construction
of the device.
6.1.1

Suitable mounting position

Mounting location
To prevent measuring errors arising from accumulation of gas bubbles in the measuring tube, avoid
the following mounting locations in the pipe:
• Highest point of a pipeline.
• Directly upstream of a free pipe outlet in a down pipe.
Installation in down pipes
Notwithstanding the above, the following installation proposal permits installation in an open down
pipe. Pipe restrictions or the use of an orifice plate with a smaller cross-section than the nominal
diameter prevent the sensor running empty while measurement is in progress.
Mounting
A0013499
17