Note) Only the Chs used are displayed on
this screen. The Ex. time is the out-
put signal hold extension time after
the completion of calibration. It is
valid only when the hold setting is
set to "ON." The Ex. time set here
is also the hold extension time at the
time of manual calibration.
Note
The gas flow for H
S measurement should be at least 5 minutes but not longer
2
than 10 minutes. Sensing element may be deteriorated if you keep the gas flow
for longer than 10 minutes.
Set Auto Cal.
key.
or
key.
ENT
34
Set flow item of
calibration gas
60 to 900 sec
ZERO
350 sec.
Ch1 Span
350 sec.
Ch2 Span
350 sec.
Ch3 Span
350 sec.
Ch4 Span
300 sec.
Ch5 Span
300 sec.
Ex. time
300 sec.
(
ENT
)