ENT
Note) Enter settings that correspond to
each range. If zirconia type is used
as O
2
of Zero (when ambient air is used),
and select the concentration listed
on the cylinder as required.
To close the setting
To close the calibration concentration value
setting process or cancel this mode midway,
ESC
press the
key.
A previous screen will return.
Setting range of values
CH
CO
, H
S
4,
2
2
The setting cannot be performed beyond the range.
Cursor for setting value
Cal. Settings
Cal. Value
CH
Ch1
0–20.00vol%
CH
0–100.0vol%
4
Ch2
0–20.00vol%
CO
0–100.0vol%
2
Ch3
0–500.0ppm
H
S
0–2000ppm
2
Ch4
0–10.00vol%
O
0–25.00vol%
2
Concentration Setting
Span gas: 1 to 105% of full scale (Full scale (FS)
is the same as each range value.)
24
ENT
(
)
Set calibration value
RANGE
ZERO
SPAN
+0000.0
0020.00
+00000
0100.00
+0000.0
0020.00
+00000
01000.0
+000.00
0500.00
+000.00
2000.00
+0000.0
0010.00
+00000
0025.00
ENT
End of Calibration
INZ-TN2ZPAF-E