Description of setting items
The alarm contact assigned the same number as the alarm is operated accordingly.
Channel:
H-Limit value:
L-Limit value:
Kind of Alarm:
ON/OFF: Enables the alarm function if set at ON, or disables it if set at OFF.
* The H-limit value cannot be set below the L-limit value, and the L-limit value cannot be set
above the H-limit value.
If it is desired to set the H-limit value below the L-limit value already stored in the memory,
reduce the L-limit value beforehand, and vice versa.
Typical on-screen display when an alarm occurs
When an H-limit alarm occurs, the "H-alarm"
message comes on in the field of relevant
Ch (component).
("L-alarm" for L-limit alarm, "HH-alarm"
for HH limit alarm, and "LL-alarm" for
LL limit alarm)
Note
After turning on power, the alarm logic trigger is inactive for 10 minutes.
INZ-TN2ZPAF-E
Channel setting targeted for issuance of alarm.
One Ch No. can be selected for multiple alarms.
Sets the high limit value (concentration) of alarm.
Sets the low limit value (concentration) of alarm.
Selects one of High limit alarm, Low limit alarm, and High limit or Low
limit alarm, HH limit alarm, and LL limit alarm.
High, HH ...... Alarm contact closes when above H-limit alarm.
Low, LL ...... Alarm contact closes when below L-limit alarm.
High or Low... Alarm contact closes when above H-limit value or
below lower limit value.
31
- - - -
C h
CH
H-alarm
4
0-20
0 0 .
C h
CO
2
0-20
.
C h
H
S
0 0 0 3
2
0-500
C h
0 0 .
O
2
0-10
vol%
vol%
ppm
vol%