6.7.1 H
S purge setting
2
Note
Changing the setting for H
knowledge and skills on gas analyzers. Do not change the default setting. If you change it,
the specified performance may not be obtained.
6.7.1.1 Explanation of H
Gas replacement
Folw time of
purge gas
Close
Solenoid valve for
sample gas
Open
Close
Solenoid valve for
purge gas
Open
measuring
position
Switching time of purge gas
Measuring time of sample gas
S purge should be performed by those who have adequate
2
S purge
2
Sample gas
measurement
Folw time of
sample gas
1 cycle
: 1 to 999s (Initial value: 60s)
: 1 to 999s (Initial value: 60s)
52
Sample gas
Gas replacement
measurement
Folw time of
Folw time of
purge gas
sample gas
INZ-TN2ZPAF-E