Fuji Electric ZPAF Instruction Manual page 50

Biomass gas analyzar
Table of Contents

Advertisement

Example
Start time
Cycle
Flow time
ON/OFF
In case where auto zero calibration is carried out at the above setting.
Sunday
12:00
Cycle
Replace-
Zero
ment
calibration
time
300sec
300sec
Flow time
(An example of "Ch1: through Ch5: enable," as given in Item 6.2.4 "Setting of auto
calibration components/range")
Setting range
Cycle
: 1 to 99 hours or 1 to 40 days (initial value 7 days)
Flow time
: 60 to 900 sec
Caution
When an auto zero calibration starts, the measurement screen automatically appears.
Any operation other than "Auto Zero Calibration Stop" (see Section 6.5.2) is not permitted
during auto zero calibration. "Auto Zero Calibration Stop" cannot be performed with
the key lock set to ON. To cancel auto zero calibration forcibly, set the key lock to OFF
and then execute "Auto Zero CalibrationStop".
If the auto calibration period and auto zero calibration period have overlapped, the auto
calibration is retained, ignoring the auto zero calibration of that period.
When the hold setting is set to ON, the hold time of auto calibration contact and
measurement value output signal is extended after calibration for gas replacement time.
Remote start
Remote start input
INZ-TN2ZPAF-E
SUN
12:00
12
hour
300
sec
ON
Monday
0:00
Gas replacement time after calibration is the same as
the flow time.
(initial value 300 sec)
Monday
12:00
: Auto zero calibration
With input (hold at least 1.5 sec.)
Without input
39